Chapter 7 ______________________________________________ TECHNICAL SPECIFICATIONS
CHAPTER 7
TECHNICAL SPECIFICATIONS
Measured Variables
Water Activity a
w
Measurement range 0...1 (-40...+60 C/-40 ...+140 F)
Accuracy 0.02(0...0.9)
0.03(0.9...1.0)
Response time (90%) at +20 C in still
oil (stainless steel filter)
10 min
Humidity sensor HUMICAP
Temperature
Measurement range -40...+100 C (-40 ...+212 F)
Typical accuracy at +20 C (+68 C) 0.1 C (0.18 F)
Typical temperature dependence of
electronics
0.005 C/C (0.003 F/F)
Temperature sensor Pt 100 RTD 1/3 Class B IEC
751