2. Connect the DMP70/GMP70/HMP70 probe to the other
connector port in the bottom of the indicator.
3. Turn on the device.
4. Check that the pressure settings of the probes (port I and II) are
the same, if you are taking measurements from the same
condition.
5. The display shows now the reading of the port I probe in the
upper row(s) and the reading of the port II in the lower row(s).
Recording Data
Recording
0605-113
Figure 18 Record and View Menu
1. Press the RECORD key. Alternatively open the MENU, select
Recording/Viewing.
2. Select Record data, press the right arrow key.
3. To change interval, select Interval and press the SET key.
4. Select the measurement interval with the up/down arrow keys.
Measurement intervals and maximum recording times are shown
in the
Table 1 on page 37.
5. Press the SELECT key.
6. To set recording duration, select Duration and press the SET
key.
7. Select the recording duration (1 min, 5 min, 15 min, 30 min, 1
h, 3 h, 12 h, 24 h, 7 days, 30 days, Memory full). Make the
selection with the arrow keys and finally press the SELECT
key.