User's Guide _______________________________________________________________________
12 ____________________________________________________________________M210498en-A
When the probe is used in a pressurized processes the sensor head should
preferebly be installed through a ball valve assembly.
Sealing thread cutting for the fitting body.
>10.5mm
fitting body
hex = 24mm
tapered thread
R1/2 ISO 7/1
ø19mm drilling
sealing with:
1. LOCTIT No 542 + activ. No 7649 (t=-55...+150 °C)
2. MEGA-PIPE EXTRA No 7188 (t=-55...+170 °C)
3. PTFE tape (t=-60...+210 °C) NOTE: the tape does not lock
the parts together. Therefore, use two fork spanners (hex 24 and
27 mm) for tightening and opening the clasp nut of the probe
parallel thread
G1/2 ISO 228/1
(BS 2779, JIS B0202)
>40mm
Process or pipe wall
CAUTION
In pressurized processed it is essential to tighten the supporting nuts
and screws very carefully to prevent loosening of the probe by the
action of pressure.
CAUTION
Take care not to damage the probe body. A damaged body makes
the probe head less tight and may prevent it from going through the
clasp nut.
Ball valve installation
The ball valve installation is preferred when connecting the probe to a
pressurized process or pipeline. Use the Vaisala DMP248BVS ball valve
set or a 1/2" ball valve assembly with a ball hole of 14 mm or more. If
you install the sensor head ( 12 mm) in a process pipe, please note that
the nominal size of the pipe must be at least 1 inch (2.54 cm). Use the
manual press tool to press the sensor head into the pressurized (< 10 bar)
process or pipeline.