Chapter 3 ________________________________________________________________ Installation
VAISALA________________________________________________________________________ 37
Cable Grounding
Ground the screen of the electrical cable properly to achieve the best
possible EMC performance.
0605-027
Figure 25 Grounding the Screen of Electrical Cable
1. Cut back outer sheath to desired length.
2. Cut back screen braiding or screen foil to dimension X (see
Figure 25 A above).
3. Push the domed cap nut (item 1 in Figure 25) and the seal insert
with contact socket of the gland (items 2 & 3) onto the cable as
shown in the diagram