OPERATING MANUAL_______________________________________________________________
58 _____________________________________________________________________ U263EN-1.4
Place the sensor head in the calibration hole of the LiCl bottle (dry
end reference) in the humidity calibrator.
Connect an ammeter/voltmeter to the analogue outputs (connector
X2). Give command l. At the first calibration point the
LED on the left flashes; adjust the first point (offset) with the
arrow switches to the value given in the calibration table (Chapter
0) and press ENT switch.
Place the sensor head in the calibration hole of the NaCl bottle
(wet end reference) in the humidity calibrator.
Wait for 10 minutes.
Check that the reading corresponds within the desired accuracy to
that given in the calibration table (Chapter 0). If not, adjust the
second point with the arrow switches to the correct value and press
ENT. At the second calibration point the second LED from the left
flashes.