Setting relay trigger points
The relays of the transmitter turn ON/OFF when the CO
2
content
reaches the predefined trigger point. There are two trigger points for
both relays. The higher limit activates the relay and the lower
deactivates it. The two limits are used to prevent the relay switching
back and forth when the measured value is very close to set point.
MF_Rx_HIGH yyy<cr>
where: x = 1 or 2 (number of the relay)
yyy = trigger point (CO
2
content in ppm)
Example of setting the higher trigger limit of the relay 1:
>MF_R1_HIGH 1200
MF_R1_HIGH=1200.0000
>
Save the settings:
MAIN_SAVE F<cr>
Give the lower trigger limit:
MF_Rx_LOW yyy<cr>
where: x= 1 or 2 (number of the relay)
yyy = trigger point (CO
2
content in ppm)
Example of setting the lower trigger limit of the relay 1:
>MF_R1_Low 900
MF_R1_LOW=900.000000
>
Save the settings:
MAIN_SAVE F<cr>
Pressure compensation
For achieving the most accurate measurements in high altitudes where
the barometric pressure is lower than in the sea level, the actual
pressure value can be set to the GMT220's software. The factory