USER'S GUIDE Calibration of Series HMD/W60/70 andHMP140 Transmitters withVaisala HUMICAP® Indicator HMI41 U218EN11

Página 2

HMI41 Calibration OptionOperating Manual 19116ZZ-U218en-1.16 1996-08-29setLoTT°C°CNumbers on the second line are blinking. They indicate the lower

Página 3 - Contents

HMI41 Calibration Option19116ZZ-U218en-1.1 Operating Manual1996-08-29 72.3.3.1 Humidity channelRHsetNumbers on the display are blinking. They indicate

Página 4

HMI41 Calibration OptionOperating Manual 19116ZZ-U218en-1.18 1996-08-29RHsetHiNumbers on the first line start to blink. They indicate the upper lim

Página 5 - 1. INTRODUCTION

HMI41 Calibration Option19116ZZ-U218en-1.1 Operating Manual1996-08-29 92.3.3.2 Temperature channelTsetNumbers on the second line are blinking. They in

Página 6 - 2 1996-08-29

HMI41 Calibration OptionOperating Manual 19116ZZ-U218en-1.110 1996-08-29setHiTNumbers on the first line start to blink. They indicate the upper lim

Página 7 - 2. OUTPUT SCALE SELECTIONS

HMI41 Calibration Option19116ZZ-U218en-1.1 Operating Manual1996-08-29 11Select the cI selection mode with buttons ▲ or ▼ and acknowledge theselection

Página 8 - 4 1996-08-29

HMI41 Calibration OptionOperating Manual 19116ZZ-U218en-1.112 1996-08-29lower righthand corner is Hi. You can now set the upper limit to correspond

Página 9 - 1996-08-29 5

HMI41 Calibration Option19116ZZ-U218en-1.1 Operating Manual1996-08-29 133. CALIBRATIONFor a successful calibration, it is essential that the probe of

Página 10 - 6 1996-08-29

HMI41 Calibration OptionOperating Manual 19116ZZ-U218en-1.114 1996-08-29 TEST CONNECTOR T GAINRHRHT OFFSETOFFSETGAINT OFFSETRHR

Página 11 - 1996-08-29 7

HMI41 Calibration Option19116ZZ-U218en-1.1 Operating Manual1996-08-29 153.2 Humidity calibrationWhen performing the humidity calibration of the HMD/W6

Página 12 - 8 1996-08-29

PUBLISHED BY Vaisala Oyj Phone (int.): +358 9 8949 1 P.O. Box 26 Fax: +358 9 8949 2227 FIN-00421 Helsinki Finland Visit our Internet pages at ht

Página 13 - 1996-08-29 9

HMI41 Calibration OptionOperating Manual 19116ZZ-U218en-1.116 1996-08-29Numbers on the first line indicate how much the RH measured by thetransmitt

Página 14 - 10 1996-08-29

HMI41 Calibration Option19116ZZ-U218en-1.1 Operating Manual1996-08-29 17If you press the HOLD button in the first T calibration mode, you will go on t

Página 15 - 1996-08-29 11

HMI41 Calibration Option19116ZZ-U218-1.1 Appendix 11996-08-26 1APPENDIX 1: QUICK REFERENCE TO SCALE SELECTIONS AND CALIBRATIONIt is recommended that

Página 16 - 12 1996-08-29

HMI41 Calibration OptionAppendix 1 19116ZZ-U218en-1.12 1996-08-262. Scale selection modesThe HMI41 automatically assumes the function selected in t

Página 17 - 3. CALIBRATION

HMI41 Calibration Option19116ZZ-U218-1.1 Appendix 11996-08-26 32.2 Current output scale selection mode (I)DISPLAY WHAT TO DO PRESS:Select the current

Página 18 - HMD70U/Y

HMI41 Calibration OptionAppendix 1 19116ZZ-U218en-1.14 1996-08-262.3 Customized voltage output selection mode (cU)DISPLAY WHAT TO DO PRESS:Select t

Página 19 - 1996-08-29 15

HMI41 Calibration Option19116ZZ-U218-1.1 Appendix 11996-08-26 52.4 Customized current output selection mode (cI)DISPLAY WHAT TO DO PRESS:Select the cu

Página 20 - 16 1996-08-29

HMI41 Calibration OptionAppendix 1 19116ZZ-U218en-1.16 1996-08-262.5 CalibrationDISPLAY WHAT TO DO PRESS:RHTWait for about 5 seconds.RH %Let the re

Página 23 - 2 1996-08-26

HMI41 Calibration Option19116ZZ-U218en-1.1 Operating Manual1996-08-29 iContents1. INTRODUCTION...

Página 24 - 1996-08-26 3

www.vaisala.com *U218EN*

Página 25 - 4 1996-08-26

HMI41 Calibration OptionOperating Manual 19116ZZ-U218en-1.1ii 1996-08-29This page intentionally left blank.

Página 26 - 1996-08-26 5

HMI41 Calibration Option19116ZZ-U218en-1.1 Operating Manual1996-08-29 11. INTRODUCTION1.1 HMI41 with calibration optionThe HMI41 can be used as a fiel

Página 27 - 6 1996-08-26

HMI41 Calibration OptionOperating Manual 19116ZZ-U218en-1.12 1996-08-29set°CIf the basic settings of the HMI41 (display units, automatic power offf

Página 28

HMI41 Calibration Option19116ZZ-U218en-1.1 Operating Manual1996-08-29 32. OUTPUT SCALE SELECTIONS2.1 Standard transmitter settingsIn Table 2.1 you wil

Página 29

HMI41 Calibration OptionOperating Manual 19116ZZ-U218en-1.14 1996-08-29Scale selection mode is entered when this text is on the display.2.3 Scale s

Página 30

HMI41 Calibration Option19116ZZ-U218en-1.1 Operating Manual1996-08-29 5setHiTT°C°CNumbers on the first line start to blink. They indicate the upper li