HMD70U/Y Transmitters
U303en-1.1 Operating Manual
1997-07-31 9
4.3 Calibration table
Temperature °C 15 20 25 30 35
°F 59 68 77 86 95
LiCl %RH *) 11.3 11.3 11.3 11.3
4...20 mA mA 5.81 5.81 5.81 5.81
0...20 mA mA 2.26 2.26 2.26 2.26
0...1 V V 0.113 0.113 0.113 0.113
0...5 V V 0.565 0.565 0.565 0.565
0...10 V V 1.13 1.13 1.13 1.13
NaCl %RH 75.6 75.5 75.3 75.1 74.9
4...20 mA mA 16.10 16.08 16.05 16.02 15.98
0...20 mA mA 15.12 15.10 15.06 15.02 14.98
0...1 V V 0.756 0.755 0.753 0.751 0.749
0...5 V V 3.780 3.775 3.765 3.755 3.745
0...10 V V 7.56 7.55 7.53 7.51 7.49
*) LiCl solution must not be used or stored in temperatures below +18 °C (+64 °F);
otherwise the equilibrium humidity of the salt solution changes permanently
Table 1 Greenspan’s calibration table with output values according to
the chosen scale
5. MAINTENANCE
5.1 Replacing the HUMICAP
180 sensor and the filter
Remove the damaged sensor and insert a new one. Handle the sensor by the
plastic socket. DO NOT TOUCH THE SENSOR PLATE. Recalibrate the
transmitter.
Replace a dirty filter (membrane or sintered) to ensure a maximum lifetime
and a fast response for the sensor. Do not try to clean the filter.