User’s Guide ______________________________________________________________________
116 _________________________________________________________________ M210906EN-C
Changing the update interval:
0TU,I=30<cr><lf>
0TU,I=30<cr><lf>
The response after the change:
0R2<cr><lf>
0R2,Ta=23.9C,Ua=26.7P<cr><lf>
Example (SDI-12, device address 0):
Changing the temperature unit to Fahrenheit:
0XTU,U=F!0<cr><lf>
In SDI-12 mode a separate enquiry (0XTU!) must be given to check the
data content.