Chapter 3 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 11
Connections
There are three different serial output interfaces available with the
PTB210 digital barometers: RS232C, RS485/232C, and
RS232C/TTL. The connections depend on the output protocol in use.
In following tables there are presented the connections of these output
interfaces.
RS232C
Table 1 Connections of the PTB210 Barometer with RS232C
Interface.
Wire color Si
nal
Gre
RX
Green TX
Blue Groun
Pin
Su
l
volta
e
5 ... 28 VDC
Brown not connecte
White not connecte
Yellow External power control
TTL level: 1 = 5 V= off
0 = 0 V= on
It is recommended to isolate the free wire ends.
RS485
Table 2 Connections of the PTB210 Barometer with
RS485/RS232C Interface
Wire color Si
nal
Gre
RX
with RS232C
Green TX
with RS232C
Blue Groun
Pin
Su
l
volta
e
8 ... 18 VDC
Brown RS485 -
White RS485 +
Yellow External power control
TTL level: 1 = off
0 = on
See also chapter "Setting the RS485 Terminating Resistor On/Off "on
page 19.
NOTE
The two output protocols cannot be used simultaneously. Connect
either RX/TX or RS485 wires. It is recommended to isolate the free
wire ends.