Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 27
SEND Outputting the reading once
In STOP mode:
SEND
In POLL mode:
SEND aa
aa = address of the transmitter when more than one transmitter is
connected to a serial bus (0...99)
Example.
>send
aw= 0.2 aw T= -47.37 'C …
If value is too long to fit to the allocated space or if there is an error in
outputting, value is displayed with stars ‘*’.
For example,
aw=*.* aw T= 31.0 'C
The output mode can be changed with command FORM.
SMODE Setting the serial interface mode
SMODE x
X= STOP/RUN/POLL
STOP MODE: Transmitter in standby for serial commands.
RUN MODE: Transmitter outputs data continuosusly.
POLL MODE: Transmitter only responds to addressed commands
>smode run
Output mode : RUN
>r
02:08:01.03 aw= 0.2 aw T= -47.13 'C …
02:08:04.21 aw= 0.2 aw T= -47.16 'C …
<ESC>
>smode stop
Output mode : STOP