Chapter 3 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 41
General Instructions for Probes with a
Cable
Mount the probes with a cable horizontally; this way, any water
condensing on the tube cannot flow onto the sensor.
0507-024
Figure 23 Horizontal Mounting of Probe
The following numbers refer to Figure 23:
1 = To be sealed.
2 = To be insulated.
3 = Insulate the cable.
4 = Let the cable hang loosely. This prevents condensed water
running to the probe along the cable.