55
Technical Data
Relative Humidity
Measurement range 0 ... 100 %RH
Typical accuracy
Temperature range 0 ... +40 °C (+32 ... +104 °F)
0 ... 90 %RH ±3 %RH
90 ... 100 %RH ±5 %RH
Temperature range -20 ... 0 °C, +40 ... +60 °C
(-4 ... +32 °F, +104 ... +140 °F)
0 ... 90 %RH ±5 %RH
90 ... 100 %RH ±7 %RH
Humidity sensor Vaisala INTERCAP
®
Stability ±2 %RH / 2 years
Response time at 20 °C (+68 °F), 90 % response 15 s
Temperature
Measurement range -20 ... +60 °C (-4 ... +140 °F)
Output scale -40 ... +60 °C (-40 ... +140 °F)
Accuracy
0 ... +40 °C (+32 ... +104 °F) ±0.4 °C (±0.72 °F)
-20 ... 0 °C, +40 ... +60 °C
(-4 ... +32 °F, +104 ... +140 °F) ±0.6 °C (±1.1 °F)
Sensor Pt1000 RTD Class F0.3 IEC 60751
General - HMD42
Supply voltage 10 ... 35 VDC (R
L
= 0 ohm)
20 ... 35 VDC (R
L
= 500 ohm)
Output signal 4 ... 20mA
General - HMD53
Supply voltage range depends on the selected output signal
DC AC
0 … 1V 10 … 35V 9 … 24V
0 … 5V 14 … 35V 12 … 24V
0 … 10V 19 … 35V 16 … 24V
WITH OPTIONAL CURRENT MODULE
0 ... 20 mA (R
L
= 0 ohm) 10 … 35V 11 … 24V
**0 … 20 mA (R
L
=500 ohm) 20 … 35V 17 … 24V
Power consumption @ 24VAC 12mA typical
**Optional temperature scales are available on request.
General
Operating temperature range
electronics -5 ... +55 °C (+23 ... +131 °F)
probe -20 ... +60 °C (-4 ... +140 °F)
Storage temperature range -40 ... +60 °C (-40 ... +140 °F)
Maximum flow speed 50 m/s
Housing material
probe stainless steel
electronics cast aluminium
Housing classification IP65 (NEMA 4)
Connections screw terminal 0.5 ... 1.5 mm
2
Cable thread-through bushing for 7 ... 10 mm
(PG9) included (18941HM)
Complies with EMC standard EN61326 and EN55022
Accessories
Armoured cable gland 10528HM
Current module (for HMD53) 18945HM
Spare Parts
Vaisala INTERCAP sensor, 1 piece 15778HM
Vaisala INTERCAP sensor, 10 pcs INTERCAPSET-10PCS
Membrane filter DRW010525
Sintered filter HM46670SP
Dimensions
Dimensions in mm
100
100
82
82
Ø 12
250
60
Ref. B211140EN-B ©Vaisala 2012
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Please contact us at
www.vaisala.com/requestinfo
www.vaisala.com
Scan the code for
more information
Humidity