User's Guide ______________________________________________________________________
44 __________________________________________________________________ M211060EN-C
Extend Analog Output Range
Use the AOVER command to allow the analog output channels to
exceed their specified range by 10%. The scaling of the quantity remains
as before; the extra range is used for additional measurement range in the
wet end.
AOVER [ON/OFF]<cr>
The following example illustrates how the analog output is affected.
Channel 1 outputs T
d
with voltage output 0 … 5 V (-40 °C … +60 °C).
After giving the AOVER ON command, the range is 0 … 5.5 V
(-40 °C … +70 °C). Note that the +60 °C T
d
point is still at 5 V.
Example:
aover on
AOVER : ON
Display Command List
Use the HELP command to display a list of the currently available
commands.
HELP<cr>
Display the Currently Active Errors
Use the ERRS command to display the currently active error codes. For
troubleshooting active errors, see Table 10 on page 57.
ERRS<cr>
Example (no activ
e errors):
errs
0000h
No errors