Vaisala DRYCAP DMT340 SERIES Guia do Utilizador Página 133

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 191
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 132
Chapter 4 _________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 131
ONLINE STATUS
Live measurement (data available): relay active (C and NO outputs are
closed)
No live data (for example: error state, sensor purge or adjustment mode):
relay released (C and NC outputs are closed)
See Figure 70 on page 131 for illustrative examples of the
FAULT/ONLINE STATUS relay output modes.
1102-040
Figure 70 FAULT/ONLINE STATUS Relay Output Modes
FAULT/ONLINE STATUS relays are usually used in conjunction with
an analog output to obtain validity information for the output value.
NOTE
If transmitter loses its power, all status
-based relays are released
similarly to the case of an instrument failure.
Vista de página 132
1 2 ... 128 129 130 131 132 133 134 135 136 137 138 ... 190 191

Comentários a estes Manuais

Sem comentários