HMP240
Appendix 7: Re-gaining M210300en
128
Example:
>regen
Regeneration started, press any key to abort
>r
S 162 RH= 20.0 %RH T= 26.5 'C heating is finished,
S 162 RH= 20.0 %RH T= 26.5 'C settling time is on and
S 162 RH= 20.0 %RH T= 26.5 'C outputs are locked
...
N 162 RH= 18.4 %RH T= 27.7 'C settling time is over and
N 162 RH= 18.5 %RH T= 27.6 'C outputs are released
N 162 RH= 18.8 %RH T= 27.5 'C
Re-gaining can not be started if automatic re-gaining (heating/cooling) is on,
the temperature of the sensor exceeds 100 °C or the security lock jumper is
off. In these cases the 'Regeneration started...' text is not output by the
transmitter but the prompt '>' appears immediately.
Example:
>regen
>r forced re-gaining failed
H 162 RH= 19.7 %RH T= 26.4 'C as automatic re-gaining
H 162 RH= 19.7 %RH T= 26.4 'C (heating) is going on
H 162 RH= 19.7 %RH T= 26.4 'C